Sample Holder - SH 100
The OmniVac SH 100 Sample Holder series provides a highly flexible mounting system for your samples which meets an outstanding range of desired experimental conditions due to their unique design.
The six electrical contacts are equipped with a thermocouple, a resistive heater and two free pins for e.g. biasing or resistance measurements. The whole sample plate can be efficiently cooled down by LN2 (down to 88 K) or LHe. Heating allows for reaching temperatures up to 2300 K. Different heating methods may be applied as e.g. direct or indirect heating, electron bombardment etc.
Many non-standard sample holders are available, e.g. quartz balance, faraday cup, charge neutralization, optical measurements in transmission etc. Modifications to meet your particular needs can easily be realized.
SH 100 Sample Holder (standard design)
Standard sample size | 1’ |
Max. sample size | 200 mm (diameter) |
Six electrical contacts | 2 for thermocouple, 2 for heating, 2 free pins (e.g. bias, resistance measurements etc.) |
Heating | up to 1300 K (indirect) resp. 2300 K (electron beam) |
Cooling | LN2 (< 100 K) or LHe (< 50 K) |
Adjustment | Internal tilt to adjust sample with measurement plane |
Design variations
Direct heating | SH 130, for e.g. flash-heating of Si/SiO2 |
Electron bombardment heating | SH 120, SH 125 (up to 2300 K) |
Interface to other sample holder systems | e.g. Omicron, SH 170, SH 175, RHK-SPM |
Heatable/coolable sample holder for optical measurements in transmission | SH 150 |
Integrated charge compensation | SH 105 |
Related Products:
- SH 100s - capableto handle Omicron and Leybold System Samples