Skip to content
OmniVac

HPM Cryo 5/6 Axes Cryo-Manipulator

Helium-cooled 5/6-axis manipulator for low-temperature angle-resolved photoemission (ARPES), open or closed cycle.

Manipulators

Description

The HPM Cryo manipulator series cools samples with helium, e.g. for low-temperature angle-resolved photoemission (ARPES). The lowest sample temperature depends on the number of independent axes and the cryostat; each quote states it. It is based on the HPM manipulator series, with an XY module on a DN 160 CF flange and a differentially pumped rotary feedthrough for full 360° Z rotation.

Features

  • Efficient shielding for low helium consumption
  • Open or closed cycle cryostat
  • Azimuthal sample rotation and/or sample tilt can be integrated
  • Sample current measurement with low dark current
  • Additional sample stage; easily accessible sample
  • Compatible with standard flag style sample holder systems
  • Full motorization and software integration for standard energy analyzers

Technical specifications

ParameterValue
CryostatHelium, open cycle or closed cycle
Cooling time RT to 10 K, open cycle< 15 min
He consumption, open cycle0.5 l/h above 10 K, 1 l/h at lowest temperature
Temperature stability, closed cycle< 0.2 K over 6 days
Indirect resistive heating< 400 K
Sample current measurement, dark current< 5×10-14 A
Mounting flangeDN 160 CF
XY stroke (full square range)±12.5 / ±25 mm
Z strokeUp to 800 mm
Resolution (manual)5 µm
Resolution (motorized)0.5 µm
Repeatability (manual)5 µm
Repeatability (motorized)1 µm
Max. speed2 mm/s
Z rotation (R1 axis)360 °
Azimuthal rotation (R2 axis)±90 °
Sample tilt (R3 axis)-15 – +45 °
Angular resolution0.1 °
Sample holdersStandard flag style sample holder systems
Included items (open cycle)Lake Shore 331S cryogenic temperature controller; Janis Research ST-400 UHV cryostat