Description
The IPES 3000 probes the density of unoccupied electronic states between the Fermi and the vacuum level, which photoemission (PES) cannot reach. Its UV bandpass photon detector is an acetone/argon filled Geiger-Müller tube with CaF₂ entrance window; the OmniVac ES 100 low-energy electron source supplies the incident electrons. Detector and electron source each mount on a DN 40 CF flange.
An indirectly heated BaO cathode gives a high beam current at low energy with a narrow energy distribution. Beam energy, beam current and spot size are set independently over wide ranges.
The Geiger-Müller detector is more sensitive to UV photons than the channeltron of the IPES 2000. The higher count rate shortens the measurement, which suits samples that degrade under the electron beam, such as organic semiconductors. It needs more setup: gas handling and adjustment of the detector pressure.
Features
- PC-driven gas mixing station for the detection and quenching gas, included
- Data acquisition software sets the electron kinetic energy and determines optimum settings for the Geiger-Müller tube and the preamplifier
Options
- Other window materials and detector configurations
Technical specifications
| Parameter | Value |
|---|---|
| Detector | GM tube with gas mixing station |
| Window | CaF₂ (standard), other materials possible |
| Gas filling | Acetone/argon (standard) |
| Resolution (overall) | 0.5 eV |
| Mounting flange | DN 40 CF |
| Data acquisition | Preamplifier and pulse counter with USB interface |
| Data processing | Software for signal counting |
| Remote control | Electron source energy control |
| ES 100 electron source — Beam energy | 1 – 200 eV |
| ES 100 electron source — Beam current | 1 nA – 20 µA |
| ES 100 electron source — Spot size (20 mm working distance) | 0.5 – 5 mm |
| ES 100 electron source — Working distance | 5 – 100, 20 recommended mm |
| ES 100 electron source — Cathode | BaO, indirectly heated |
| ES 100 electron source — Mounting flange | DN 40 CF |
| ES 100 electron source — Power supply | All necessary voltages to drive the electron source |
| ES 100 electron source — Interface | USB |