Description
The SHR 100 receives sample holders of the SH 100 Sample Holder series. Pneumatically driven mechanics contact six electrical connections, e.g. thermocouple, heater and 2× bias, and the cooling finger with its sapphire sphere. All contacts, heating and cooling stay connected during sample positioning and continuous azimuthal rotation.
The manipulator it sits on provides X, Y, Z and polar rotation; azimuthal rotation and tilt can be added in the station itself.
With the matching SH 100 holder, the station supports indirect heating through a Cu or Mo plate, electron beam or direct heating, and LN₂ or LHe cooling. The reachable temperature depends on the holder.
Technical specifications
| Parameter | Value |
|---|---|
| Sample holders | SH 100 series (1″); Prevac 1″ holders also fit |
| Electrical contacts | 6: 2 thermocouple, 2 resistive heater, 2 free pins |
| Thermocouple | Type K |
| Max. heating or sample current | 15 A |
| Max. temperature | Depends on the sample holder |
| Max. temperature, indirect heating through Cu plate | 900 K |
| Max. temperature, indirect heating through Mo plate | 1300 K |
| Max. temperature, electron beam or direct heating | 1500 K |
| Max. temperature, electron beam heating with reduced cooling | 2300 K |
| Cooling | Cooling finger with sapphire sphere; LN₂ or LHe |
| Temperature, LN₂ cooling | < 100 K |
| Sample holder locking | Pneumatic |
| Rotation | Continuous azimuthal; contacts, heating and cooling maintained |
| Axes | Azimuth and tilt in the station; X, Y, Z and polar rotation from the manipulator |
| Materials | Non-magnetic |
| Mounting | Fixed room coordinates or rotatable HPM XYZ manipulator |